Image Attributes


Attribute Name Tag Type VR VM Descriptiom Creator Unit
General Image Attributes        
AcquisitionDate 0008,0022 3 DA 1 A number identifying the single continuous gathering of data over a period of time which resulted in this image acquisition  
ContentDate 0008,0023 2C DA 1 The date the image data creation started. Required if image is part of a series in which the images are temporally related acquisition  
AcquisitionTime 0008,0032 3 TM 1 The time the acquisition of data that resulted in this image started acquisition  
ContentTime 0008,0033 2C TM 1 The time the image pixrl data creation started. Required if image is part of a series in which the images are temporally related acquisition  
Acquisition Datetime 0008,002A 3 DT 1 The date and time that the acquisition of dada that result in this image started    
DerivationDescription 0008,2111 3 ST 1 A text description of how this image was derived image creator  
AcquisitionNumber 0020,0012 2 IS 1 Image identification characteristics acquisition  
InstanceNumber 0020,0013 2 IS 1 A number that identifies this image acquisition  
ImageComments 0020,4000 3 LT 1 User-defined comments about the image acquisition user interface  
ImageInAcquisition 0020,1002 3 IS 1 Number of images that resulted from this acquisition of data    
ReferencedImageSequence 0008,1140 3 link 1 A sequence which provides reference to a set of Image SOP Class/Instance identifying other images significantly related to this image (localizer images) acquisition  
>Referenced SOP Class UID 0008,1150 1C UI 1 Uniquely identifies the referenced SOP Class acquisition  
>Referenced SOP Instance UID 0008,1155 1C UI 1 Uniquely identifies the referenced SOP Instance acquisition  
> Referenced Frame Number 0008,1160 3 IS 1-n References one or more image frames of a Multi-frame Image SOP Instance, identifying which frames are siginificantly related to this image acquisition  
SourceImageSequence 0008,2112 3 link 1 A sequence which identifies the set of Image SOP Class/Instance pairs of the images which were use to derive this image acquisition  
>Referenced SOP Class UID 0008,1150 1C UI 1 Uniquely identifies the referenced SOP Class acquisition  
>Referenced SOP Instance UID 0008,1155 1C UI 1 Uniquely identifies the referenced SOP Instance acquisition  
> Referenced Frame Number 0008,1160 3 IS 1-n References one or more image frames of a Multi-frame Image SOP Instance, identifying which frames are siginificantly related to this image acquisition  
ImageType 0008,0008 3 CS 1-n Image identification characteristics acquisition  
Lossy Image Compression 0028,0300 3 CS 1 Specifies whether an image has undergone lossy compression. Enumerated values: 00 = image has not been subjected to lossy compression. 01 = image has been subjected to lossy compression image creator  
Quality Control Image 0008,0005 1C CS 1 Indicates whether or not this image is a quality control or phantom image. Enumerated values: YES, NO image creator  
PatientOrientation 0020,0020 2 CS k Patient direction of the rows and columns of the image. Not required for MR images    
MR Image Attributes        
ImageType 0008,0008 1 CS 1-n Image identification characteristics acquisition  
Samples per Pixel 0028,0002 1 US 1 Number of samples (planes) in this image acquisition  
PhotometricInterpretation 0028,0004 1 CS 1 Specifies the intended interpretation of the pixel data acquisition  
BitsAllocated 0028,0100 1 US 1 Number of bits allocated for each pixel sample. Each sample shall have the same number of bits allocated acquisition  
Scanning Sequence 0018,0020 1 CS 1-n Description of the type of data taken. Enumerated values: SE = Spin Echo; IR = Inversion Recovery; GR = Gradient Recalled; EP = Echo Planar; RM = Research Mode acquisition  
SequenceVariant 0018,0021 1 CS 1-n Variant of the Scanning Sequence acquisition  
ScanOptions 0018,0022 2 CS 1-n Parameters of scanning sequence acquisition  
MRAcquisitionType 0018,0023 2 CS 1 Identification of date encoding scheme. Enumerated Values: 2D = frequency x phase 3D = frequency x phase x phase acquisition  
               
SequenceName 0018,0024 3 SH 1 User defined name for the Scanning Sequence and Sequence Varaint combination acquisition  
AngioFlag 0018,0025 3 CS 1 Angio image indicator. Primary image for angio processing. Enumerated values: Y = image is Angio N = image is not Angio acquisition  
RepetitionTime 0018,0080 2C DS 1 The period of time in msec between the beginning of a pulse sequence and the beginning of a succeeding (essentially identical) pulse sequence. Required except when Scanning Sequence is EP and Sequnece Variant is not segmented k-space acquisition msec
EchoTime 0018,0081 2 DS 1 Time in msec between the middle of the excitation pulse and the peak of the echo produced. In the case of segmented k-space, the TE(eff) is the time between the middle of the excitation pulse to the peak of the echo that is used to cover the center of k-space) acquisition msec
InversionTime 0018,0082 2C DS 1 Time in msec after the middle of inverting RF pulse to middle of excitation pulse to detect the amount of longitudinal magnetization. Required if Scanning Sequence has value of IR acquisition msec
NumberOfAverages 0018,0083 3 DS 1 Number of times a given pulse sequence is repeated before any parameter has changed acquisition  
ImagingFrequency 0018,0084 3 DS 1 Precession frquency in MHz of the nucleus being addressed acquisition MHz
ImagedNucleus 0018,0085 3 SH 1 Nucleus that is resonant at the imaging frequency acquisition  
EchoNumbers 0018,0086 3 IS 1-n The echo number used in generating this image. In the case of segmented k-space, it is the effective Echo Number acquisition  
MagneticFieldStrength 0018,0087 3 DS 1 Nominal Field strength of MR magnet, in Tesla acquisition Tesla
SpacingBetweenSlices 0018,0088 3 DS 1 Spacing between slices, in mm. The spacing is measured from the center-to-center of each slice acquisition mm
NumberOfPhaseEncodingSteps 0018,0089 3 IS 1 Total number of lines in k-space in the y-direction collecting during acquisition acquisition  
EchoTrainLength 0018,0091 2 IS 1 Number of lines in k-space acquired per excitation per image acquisition  
PercentSampling 0018,0093 3 DS 1 Fraction of acquisition matrix lines acquired, expressed as a percent acquisition  
PercentPhaseFieldOfView 0018,0094 3 DS 1 Ration of field of view dimension in phase direction to field of view dimension in frequency direction, expressed as a percent acquisition  
PixelBandwidth 0018,0095 3 DS 1 Reciprocal of the total sampling period, in hertz per pixel acquisition hertz/pixel
BeatRejectionFlag 0018,1080 3 CS 1 Beat length sorting has been applied. Enumerated values: Y = yes N = no acquisition  
Low R-R Value 0018,1081 3 IS 1 R-R interval low limit for beat rejection, in msec acquisition msec
High R-R Value 0018,1082 3 IS 1 R-R interval high limit for beat rejection, in msec acquisition msec
TriggerTime 0018,1060 2C DS 1 Time, in msec, between peak of the R wave and the peak of the echo produced. In the case of segmented k-space, the TE(eff) is the time between the peak of the echo that is used to cover the center of k-space. Required for Scan Options which include heart gating acquisition msec
NominalInterval 0018,1062 3 IS 1 Average R-R interval used for the scans, in msec acquisition msec
IntervalsAcquired 0018,1083 3 IS 1 Number of R-R intervals acquired acquisition  
IntervalsRejected 0018,1083 3 IS 1 Number of R-R intervals rejected acquisition  
PVC Rejection 0018,1085 3 LO 1 Description of type of PVC rejection criteria used acquisition  
Skip Beats 0018,1086 3 IS 1 Number of beats skipped after a detected arrhythmia acquisition  
HeartRate 0018,1088 3 IS 1 Beats per minute acquisition 1/min
CardiacNumberOfImages 0018,1090 3 IS 1 Number of imagesb per cardiac cycle acquisition  
TriggerWindow 0018,1094 3 IS 1 Percent of R-R interval, based on Heart Rate (0018,1088), prescriped as a window for a valid/usable trigger    
ReconstructionDiameter 0018,1100 3 DS 1 Diameter in mm of the region from within which data were used in creating the reconstruction of the image. Data may exist outside this region and portions of the patient may exist outside the region acquisition  
Receive Coil 0018,1250 3 SH 1 Received coil used acquisition  
TransmittingCoil 0018,1251 3 SH 1 Transmitted coil used acquisition  
AcquisitionMatrix 0018,1310 3 US 4 Dimension of acquired frequency/phase data before reconstruction. Multi-valued: frequency rows\frequency columns\phase rows\phase columns acquisition  
PhaseEncodingDirection 0018,1312 3 CS 1 The axis of phase encoding with respect to the image. Enumerated Values: ROW = phase encoded in rows COL = phase encoded in columns acquisition  
FlipAngle 0018,1314 3 DS 1 Steady state angle in degrees to which the magnetic vector is flipped from the magnetic vector to the primary field acquisition degree
VariableFlipAngleFlag 0018,1315 3 CS 1 Flip angle variation applied during image acquisition. Enumerated Values: Y = yes N = no acquisition  
SAR 0018,1316 3 DS 1 Calculated whole body specific absortion rate in watts/kilogram acquisition Watt/kilogram
dbdt 0018,1318 3 DS 1 The rate of change of the gradient coil magnetic flux density with time (T/s) acquisition Tesla/sec
Contrast Bolus Attributes        
ContrastBolusAgent 0018,0010 2 LO 1 Contrast or bolus agent acquisition user interface  
ContrastBolusStartTime 0018,1042 3 TM 1 Time of start of injection acquisition user interface  
ContrastBolusStopTime 0018,1043 3 TM 1 Time of end of contrast injection acquisition user interface  
ContrastBolusTotalDose 0018,1044 3 DS 1 Total amount in milliliters of the indiluted contrast agent acquisition user interface ml
Contrast Flow Rate(s) 0018,1046 3 DS 1-n Rate(s) of injection(s) in millilitres/sec acquisition user interface ml/sec
ContrastBolusVolume 0018,1041 3 DS 1 Volume injected in milliliters of diluted contrast agent acquisition user interface  
ContrastFlowDurations 0018,1047 3 DS 1-n Duration(s) of injection(s) in seconds. Each Contrast Flow Duration value shall correspond to a value of Contrast Flow Rate (0018,1046) acquisition user interface sec
               
Contrast/Bolus Ingredient 0018,1048 3 CS 1 Active ingredient of agent, Defined Terms: IODINE, GADOLIMIUM, CARBON DIOXIDE, BARIUM acquisition user interface  
Contrast/Bolus Ingredient Concentration 0018,1049 3 DS 1 Milligrams of active ingredient per milliliter of (diluted) agent acquisition user interface mg/ml
Image Pixel Attributes        
SamplesPerPixel 0028,0002 1 US 1 Number of samples (planes) in this image acquisition  
PhotometricInterpretation 0028,0004 1 CS 1 Specifies the intended interpretation of the pixel data acquisition  
Rows 0028,0010 1 US 1 Number of rows in the image acquisition  
Columns 0028,0010 1 US 1 Number of columns in the image acquisition  
PixelAspectRatio 0028,0034 1C IS 2 Ratio of the vertical size and horizontal size of the pixels in the image specified by a pair of integer values where the first value is the vertical pixel size, and the second value is the horizontal pixel size. Required if the aspect ratio is not 1/1 and the Image Plane Module is not applicable to this Image acquisition  
BitsAllocated 0028,0100 1 US 1 Number of bits allocated for each pixel sample. Each sample shall have the same number of bits allocated acquisition  
BitsStored 0028,0101 1 US 1 Number of bits stored for each pixel sample. Each sample shall have the same number of bits stored acquisition  
HighBit 0028,0102 1 US 1 Most significant bit for pixel sample data. Each sample shall have the same high bit acquisition  
PixelRepresentation 0028,0103 1 US 1 Data representation of the pixel samples. Each sample shall have the same pixel representation. Enumerated values: 0000H = unsigned integer. 0001H = 2's complement. acquisition  
PixelData 7FE0,0010 1 OW/OB 1 A data stream of the pixel samples which comprises the image acquisition  
Smallest Image Pixel Value 0028,0106 3 US/SS 1 The minimum actual pixel value encountered in this image acquisition  
Largest Image Pixel Value 0028.0107 3 US/SS 1 The maximum actual pixel value encountered in this image acquisition  
Image Plane Attributes        
SliceThickness 0018,0050 2 DS 1 Nominal slice thickness, in mm acquisition mm
ImageOrientationPatient 0020,0037 1 DS 6 The direction cosines of the first row and the first column with respect to the patient acquisition  
ImagePositionPatient 0020,0032 1 DS 3 The x,y and z coordinates of the upper left and hand corner (first pixel transmitted) of the image, in mm acquisition mm
SliceLocation 0020,1041 3 DS 1 Relative position of exposure expressed in mm acquisition mm
PixelSpacing 0028,0030 1 DS 2 Physical distance in the patient between the center of each pixel, specified by a numeric pair-adjacent row spacing (delimiter) adjacent column spacing in mm acquisition mm
VOI Lookup Table        
WindowCenter 0028,1050 3 DS 1-n Window center acquisition/application  
WindowWidth 0028,1050 1C DS 1-n Window width acquisition/application  
WindowCenterWidthExplanation 0028,1055 3 LO 1-n Free form explanation of the meaning of the Window Center and Width. Multiple values correspond to multiple Window Center and Width values acquisition/application  
Equipment Attributes              
Manufacturer 0008,0070 2 LO 1 Manufacturer of the equipment that produced the digital images acquisition/application  
InstitutionName 0008,0080 3 LO 1 Institution where the equipment is located that produced the digital images acquisition/application  
InstitutionAddress 0008,0081 3 ST 1 Mailing address of the institution where the equipment is located that produced the digital images acquisition/application  
ManufacturersModelName 0008,1090 3 LO 1 Manufacturer's model number of the equipment that produced the digital images acquisition/application  
DeviceSerialNumber 0018,1000 3 LO 1 Manufacturer's seriel number of the equipment that produced the digital images acquisition/application  
SoftwareVersions 0018,1020 3 LO 1-n Manufacturer's designation of software of the equipment that produced the digital images acquisition/application  
Overlay Idendification Attributes              
OverlayRows 60xx,0010 1 US 1 Number of Rows in Overlay post-processing applications  
OverlayColumns 60xx,0011 1 US 1 Number of Columns in Overlay post-processing applications  
OverlayType 60xx,0040 1 CS 1 Indicates whether this overlay represents a region of interest or other graphics post-processing applications  
Overlay Origin 60xx,0050 1 US 1 Location of first overlay point with respect to pixel in the image, given as row\column post-processing applications  
OverlayBitsAllocated 60xx,0100 1 US 1 Number of bits allocated in the overlay post-processing applications  
OverlayBitPosition 60xx,0102 1 US 1 Bit in which overlay is stored post-processing applications  
Overlay Data 60xx,3000 1C OB/OW 1 Overlay pixel data post-processing applications  
MEDCOM Header              
HeaderType 0029,xx08 1C CS 1 Medcom header characteristics. Defined Terms: MEDCOM 1 acquisition  
HeaderVersion 0029,xx09 2C LO 1 Version of Medcom header acquisition  
HeaderInfo 0029,xx10 3 OB 1 Manufacturer header info acquisition  
HistoryInfo 0029,xx20 3   1 Patient registration history patient registration  
MR and SC Private Attributes              
EchoLinePosition stream 1 IS 1 Fourier line position with the maximal echo for the performed acquisition acquisition  
EchoColumnPosition stream 1 IS 1 Echo column position for the performed acquisition acquisition  
EchoPartitionPosition stream 1 IS 1 Echo partition position for the performed acquisition acquisition  
UsedChannelMask stream 1 UL 1 8 bit mask of the used receiver channels for the performed acquisition. Example: channel 0: 00000001 channel 3: 00000111 · acquisition  
Actual3DImaPartNumber stream 1 IS 1 Number of a 3D partitions beginning with 0 acquisition  
ICE_Dims stream 1 LO 1 The 9 used ICE object dimensions of the performed acquisition. Combined/unset dimensions will be marked with "X". E.g.: X_2_1_1_1_1_2_1_1 acquisition  
B_value stream   IS 1 Diffusion effect in s/mm*mm of the ICE program for the performed acquisition acquisition sec/(mm*mm)
Filter1 stream 1 IS   Context vision filter acquisition  
Filter2 stream 1 IS   not used acquisition  
ProtocolSliceNumber stream 1 IS 1 Number of the slice beginning with 0 acquisition  
RealDwellTime stream 1 IS 1 The time in ns between the beginning of sampling one data point and the beginning of sampling of next data point in the acquired signal. This means the dwell time is the sampling rate during digital conversion of an acquired signal acquisition nanosec
PixelFile stream 1 UN 1 Used raw data file for the performed acquisition acquisition  
PixelFileName stream 1 UN 1 Used raw data file name for the performed acquisition    
SliceMeasurementDuration stream 1 DS 1 Time duration between two slices of the performed acquisition acquisition msec
AcquisitionMatrixText stream 1 SH 1 Used acquisition matrix description acquisition  
SequenceMask stream 1 UL 1 Parameters used for acquisition, e.g. door open, interpolation, raw filter, Siemens seqence .... acquisition  
MeasuredFourierLines stream 1 IS 1 Number of performed fourier lines acquisition  
FlowEncodingDirection stream   IS 1 Flow encoding direction acquisition  
FlowVenc stream   FD 1 Flow Quant attribute acquisition  
PhaseEncodingDirectionPositive stream   IS 1 Phase encoding direction: 0 = negative; 1 = positive acquisition  
NumberOfImagesInMosaic stream   US 1 Number of slices in a mosaic image acquisition  
DiffusionGradientDirection stream   FD 3 Diffusion in gradient direction acquisition  
ImageGroup stream   US 1 Group of images acquisition  
SliceNormalVector stream   FD 3 X,y and z normal vector of the slices acquisition  
DiffusionDirection stream   CS 1 Diffusion direction acquisition  
TimeAfterStart stream   DS 1 Time delay after start of measurment acquisition  
FlipAngle stream   DS 1 Flip angle for SC images acquisition degree
SequenceName stream   SH 1 Sequence name for SC images acquisition  
RepetitionTime stream   DS 1 Repetition time for SC images acquisition msec
EchoTime stream   DS 1 Echo time for SC images acquisition msec
NumberOfAverages stream   DS n Number of averages for SC images acquisition  
MR NonImage (Raw Data, Spectra) Attributes              
NonimageType 0029,xx08 1 CS 1 Data identification characteristics. Defined terms: RAW DATA NUM 4; SPEC NUM4 acquisition  
NonimageVersion 0029,xx09 3 LO 1 Version of Non-image data acquisition  
NonimageInfo 0029,xx10 3 OB 1 Description of Non-image data acquisition  
NonimageData 7FE1,xx10 2 OB 1 Binary data stream acquisition  
ImageType 0008,0008 3 CS 1-n Image identification characteristics acquisition  
AcquisitionDate 0008,0022 3 DA 1 The date the acquisition of data started acquisition  
AcquisitionTime 0008,0032 3 TM 1 The time the acquisition of data started acquisition  
DerivationDescription 0008,2111 3 ST 1 A text description of how this data set was derived acquisition  
AcquisitionNumber 0020,0012 2 IS 1 A number identifying the gathering of data over a period of time which resulted in this data set acquisition  
ImageNumber stream   IS 1 A number that identifies this image acquisition  
ImageComments stream   LT 1 User-defined comments about the image acquisition  
ReferencedImageSequence stream   UI n A sequence which provides reference to a set of Image SOP Class/Instance identifying other images significantly related to this image (localizer images) acquisition  
PatientOrientation stream   CS 1 Patient direction of the rows and columns of the image. Not required for MR images acquisition  
ScanningSequence stream   CS n Description of the type of data taken. Enumerated values: SE = Spin Echo; IR = Inversion Recovery; GR = Gradient Recalled; EP = Echo Planar; RM = Research Mode acquisition  
SequenceName stream   SH 1 User defined name for the Scanning Sequence and Sequence Varaint combination acquisition  
RepetitionTime stream   DS 1 The period of time in msec between the beginning of a pulse sequence and the beginning of a succeeding (essentially identical) pulse sequence. Required except when Scanning Sequence is EP and Sequnece Variant is not segmented k-space acquisition  
EchoTime stream   DS 1 Time in msec between the middle of the excitation pulse and the peak of the echo produced. In the case of segmented k-space, the TE(eff) is the time between the middle of the excitation pulse to the peak of the echo that is used to cover the center of k-space) acquisition  
InversionTime stream   DS 1 Time in msec after the middle of inverting RF pulse to middle of excitation pulse to detect the amount of longitudinal magnetization. Required if Scanning Sequence has value of IR acquisition  
NumberOfAverages stream   DS 1 Number of averages acquisition  
ImagingFrequency stream   DS 1 Precession frquency in MHz of the nucleus being addressed acquisition  
ImagedNucleus stream   SH 1 Nucleus that is resonant at the imaging frequency acquisition  
EchoNumbers stream   IS 1 The echo number used in generating this image. In the case of segmented k-space, it is the effective Echo Number acquisition  
MagneticFieldStrength stream   DS 1 Nominal Field strength of MR magnet, in Tesla acquisition Tesla
NumberOfPhaseEncodingSteps stream   IS 1 Total number of lines in k-space in the y-direction collecting during acquisition acquisition  
EchoTrainLength stream   IS 1 Number of lines in k-space acquired per excitation per image acquisition  
PercentSampling stream   DS 1 Fraction of acquisition matrix lines acquired, expressed as a percent acquisition  
PercentPhaseFieldOfView stream   DS 1 Ration of field of view dimension in phase direction to field of view dimension in frequency direction, expressed as a percent acquisition  
TriggerTime stream   DS 1 Time, in msec, between peak of the R wave and the peak of the echo produced. In the case of segmented k-space, the TE(eff) is the time between the peak of the echo that is used to cover the center of k-space. Required for Scan Options which include heart gating acquisition  
ReceivingCoil stream   SH 1 Received coil used acquisition  
TransmittingColi stream   SH 1 Transmitted coil used acquisition  
AcquisitionMatrixText stream   US 4 Dimension of acquired frequency/phase data before reconstruction. Multi-valued: frequency rows\frequency columns\phase rows\phase columns acquisition  
PhaseEncodingDirection stream   CS 1 The axis of phase encoding with respect to the image. Enumerated Values: ROW = phase encoded in rows COL = phase encoded in columns acquisition  
FlipAngle stream   DS 1 Steady state angle in degrees to which the magnetic vector is flipped from the magnetic vector to the primary field acquisition Degree
VariableFlipAngleFlag stream   CS 1 Flip angle variation applied during image acquisition. Enumerated Values: Y = yes N = no acquisition  
SAR stream   DS 1 Calculated whole body specific absortion rate in watts/kilogram acquisition Watt/kilogram
dBdt stream   DS 3 The rate of change of the gradient coil magnetic flux density with time (T/s) acquisition Tesla/sec
Rows stream   US 1 Number of rows in the image acquisition mm
Columns stream   US 1 Number of columns in the image acquisition mm
SliceThickness stream   DS 1 Nominal slice thickness, in mm acquisition mm
ImagePositionPatient stream   DS 3 The direction cosines of the first row and the first column with respect to the patient acquisition  
ImageOrientationPatient stream   DS 6 The x,y and z coordinates of the upper left and hand corner (first pixel transmitted) of the image, in mm. acquisition mm
SliceLocation stream   DS 1 Relative position of exposure expressed in mm acquisition mm
EchoLinePosition stream   IS 1 Fourier line position with the maximal echo for the performed acquisition acquisition  
EchoColumnPosition stream   IS 1 Echo column position for the performed acquisition acquisition  
EchoPartitionPosition stream   IS 1 Echo partition position for the performed acquisition acquisition  
Actual3DImaPartNumber stream   IS 1 Number of a 3D partitions beginning with 0 acquisition  
RealDwellTime stream   IS 1 The time in ns between the beginning of sampling one data point and the beginning of sampling of next data point in the acquired signal. This means the dwell time is the sampling rate during digital conversion of an acquired signal acquisition  
ProtocolSliceNumber stream   UN 1 Number of the slice beginning with 0 acquisition  
PixelFile stream   UN 1 Used raw data file for the performed acquisition acquisition  
PixelFileName stream   LO 1 Used raw data file name for the performed acquisition acquisition  
ICE_Dims stream   DS 1 The 9 used ICE object dimensions of the performed acquisition. Combined/unset dimensions will be marked with "X". E.g.: X_2_1_1_1_1_2_1_1 acquisition  
PixelSpacing stream   DS 2 Physical distance in the patient between the center of each pixel, specified by a numeric pair-adjacent row spacing (delimiter) adjacent column spacing in mm acquisition mm
SourceImageSequence stream   UI n A sequence which identifies the set of Image SOP Class/Instance pairs of the images which were use to derive this image acquisition  
PixelBandwidth stream   DS 1 Reciprocal of the total sampling period, in hertz per pixel acquisition  
SliceMeasurementDuration stream   DS 1 Time duration between two slices of the performed acquisition acquisition  
SequenceMask stream   UL 1 Parameters used for acquisition, e.g. door open, interpolation, raw filter, Siemens seqence .· acquisition  
AcquisitionMatrixText stream   SH 1 Used acquisition matrix description acquisition  
MeasuredFourierLines stream   IS 1 Number of performed fourier lines acquisition  
MR Protocol              
ulVersion stream 3 UL 1 Protocol version acquisition  
tSequenceFileName stream 3 ST 1 Sequence file name for actual measurement protocol acquisition  
tProtocolName stream 3 ST 1 Name of actual measurement protocol acquisition  
tReferenceImage0 stream 3 LO 1 Referenced image acquisition  
tReferenceImage1 stream 3 LO 1 Referenced image acquisition  
tReferenceImage2 stream 3 LO 1 Referenced image acquisition  
lScanRegionPosSag stream 3 FD 1 Desired table position in series block coordinate system acquisition  
lScanRegionPosCor stream 3 FD 1 Desired table position in series block coordinate system acquisition  
lScanRegionPosTra stream 3 FD 1 Desired table position in series block coordinate system acquisition  
ucScanRegionPosValid stream 3 SH 1 Valid flag for desired table position in series block coordinate system acquisition  
aucSpare stream 3 SH 1 Reserved acquisition  
lScanRegionDelta stream 3 SL 1 Scan Region Position/Move [mm] acquisition mm
sProtConsistencyInfo.tBaselineString stream 3 LO 1 Baseline consistence info acquisition  
sProtConsistencyInfo.tGradCoilName stream 3 SH 1 Coil name consistence info acquisition  
sProtConsistencyInfo.tGradAmplifierName stream 3 LO 1 Gradient amplifier consistence info acquisition  
sProtConsistencyInfo.flNominalB0 stream 3 FD 1 Nominal Bo compensation consistence acquisition  
sGRADSPEC.sEddyCompensationX.aflAmplitude stream 3 FD 5 Eddy compensation x amplitude gradient system specification acquisition  
sGRADSPEC.sEddyCompensationX.aflTimeConstant stream 3 FD 5 Eddy compensation x time parameter gradient system specification acquisition  
sGRADSPEC.sEddyCompensationY.aflAmplitude stream 3 FD 5 Eddy compensation y amplitude gradient system specification acquisition  
sGRADSPEC.sEddyCompensationY.aflTimeConstant stream 3 FD 5 Eddy compensation y time parameter gradient system specification acquisition  
sGRADSPEC.sEddyCompensationZ.aflAmplitude stream 3 FD 5 Eddy compensation z amplitude gradient system specification acquisition  
sGRADSPEC.sEddyCompensationZ.aflTimeConstant stream 3 FD 5 Eddy compensation z time parameter gradient system specification acquisition  
sGRADSPEC.bB0CompensationValid stream 3 SL 1 B0 compensation gradient system specification valid flag acquisition  
sGRADSPEC.sCrossTermCompensationXY.aflAmplitude stream 3 FD 5 Crossterm compensation xy amplitude gradient system specification acquisition  
sGRADSPEC.sCrossTermCompensationXY.aflTimeConstant stream 3 FD 5 Crossterm compensation xy time parameter gradient system specification acquisition  
sGRADSPEC.sCrossTermCompensationXZ.aflAmplitude stream 3 FD 5 Crossterm compensation xz amplitude gradient system specification acquisition  
sGRADSPEC.sCrossTermCompensationXZ.aflTimeConstant stream 3 FD 5 Crossterm compensation xz time parameter gradient system specification acquisition  
sGRADSPEC.sCrossTermCompensationYX.aflAmplitude stream 3 FD 5 Crossterm compensation xz amplitude gradient system specification acquisition  
sGRADSPEC.sCrossTermCompensationYX.aflTimeConstant stream 3 FD 5 Crossterm compensation yx time parameter gradient system specification acquisition  
sGRADSPEC.sCrossTermCompensationYZ.aflAmplitude stream 3 FD 5 Crossterm compensation yz amplitude gradient system specification acquisition  
sGRADSPEC.sCrossTermCompensationYZ.aflTimeConstant stream 3 FD 5 Crossterm compensation yz time parameter gradient system specification acquisition  
sGRADSPEC.sCrossTermCompensationZX.aflAmplitude stream 3 FD 5 Crossterm compensation zx amplitude gradient system specification acquisition  
sGRADSPEC.sCrossTermCompensationZX.aflTimeConstant stream 3 FD 5 Crossterm compensation zx time parameter gradient system specification acquisition  
sGRADSPEC.sCrossTermCompensationZY.aflAmplitude stream 3 FD 5 Crossterm compensation zx amplitude gradient system specification acquisition  
sGRADSPEC.sCrossTermCompensationZY.aflTimeConstant stream 3 FD 5 Crossterm compensation zy time parameter gradient system specification acquisition  
sGRADSPEC.bCrossTermCompensationValid stream 3 SL 1 Crossterm compensation gradient system specification valid flag acquisition  
sGRADSPEC.lOffsetX stream 3 SL 1 Gradient offset x direction [bit pattern] acquisition  
sGRADSPEC.lOffsetY stream 3 SL 1 Gradient offset y direction [bit pattern] acquisition  
sGRADSPEC.lOffsetZ stream 3 SL 1 Gradient offset z direction [bit pattern] acquisition  
sGRADSPEC.bOffsetValid stream 3 SL 1 Gradient offsets valid flag acquisition  
sGRADSPEC.lDelayX stream 3 SL 1 Gradient delay x direction acquisition  
sGRADSPEC.lDelayY stream 3 SL 1 Gradient delay y direction acquisition  
sGRADSPEC.lDelayY stream 3 SL 1 Gradient delay z direction acquisition  
sGRADSPEC.bDelayValid stream 3 SL 1 Gradient delay valid flag acquisition  
sGRADSPEC.flSensitivityX stream 3 FD 1 Gradient sensitivity x direction [mT/m] acquisition mT/m
sGRADSPEC.flSensitivityY stream 3 FD 1 Gradient sensitivity y direction [mT/m] acquisition mT/m
sGRADSPEC.flSensitivityZ stream 3 FD 1 Gradient sensitivity z direction [mT/m] acquisition mT/m
sGRADSPEC.bSensitivityValid stream 3 SL 1 Gradient sensitivity valid flag acquisition  
sGRADSPEC.flGSWDMinRiseTime stream 3 FD 1 Minimum gradient rise time for mode GRAD_GSWD_RISETIME acquisition mA
sGRADSPEC.alShimCurrent stream 3 SL 15 Shim current parameter [mA] acquisition  
sGRADSPEC.bShimCurrentValid stream 3 SL 1 Shim current parameter valid flag acquisition  
sGRADSPEC.ucMode stream 3 SH 1 Gradient mode: fast, normal, whisper acquisition  
sGRADSPEC.ucSpare0 stream 3 SH 1 Reserved acquisition  
sGRADSPEC.ucSpare1 stream 3 SH 1 Reserved acquisition  
sGRADSPEC.ucSpare2 stream 3 SH 1 Reserved acquisition  
sTXSPEC.asNucleusInfo[0].tNucleus stream 3 SH 1 Transmitter system nucleus acquisition  
sTXSPEC.asNucleusInfo[0].lFrequency stream 3 SL 1 Transmitter system frequency [Hz] acquisition Hz
sTXSPEC.asNucleusInfo[0].bFrequencyValid stream 3 SL 1 Frequency valid flag acquisition  
sTXSPEC.asNucleusInfo[0].lDeltaFrequency stream 3 SL 1 Offset from center frequency (lFrequency) acquisition  
sTXSPEC.asNucleusInfo[0].flReferenceAmplitude stream 3 FD 1 Transmitter reference amplitude [V] acquisition V
sTXSPEC.asNucleusInfo[0].bReferenceAmplitudeValid stream 3 SL 1 Reference amplitude valid flag acquisition  
sTXSPEC.asNucleusInfo[0].flAmplitudeCorrection stream 3 FD 1 Transmitter amplitude correction factor, e.g. used for water suppression acquisition  
sTXSPEC.asNucleusInfo[0].bAmplitudeCorrectionValid stream 3 SL 1 Amplitude correction valid flag acquisition  


Last Updated on 07.02.2002